Designed for steam sterilization processes of single use medical devices that require strong mechanical protection, with a target event related sterility maintenance of up to 3 years, Polybond-S 112 gsm associates medium to high seal strength levels to efficiently withstand vacuum cycles during sterilization, controlled maximum pore size for highest possible safety in sterility maintenance after sterilization, and very good mechanical performance in dry and wet stages thanks to polymer reinforcement of web material to maintain physical protection of your medical device.
Polybond-S 112 gsm has been developed to seal to PA/PP films for the packaging of single use medical devices such as pre-filled syringes as well as the segment of industrial steam sterilization businesses where Tyvek packaging material is currently being used
Material description
POROUS WEB | |||
Product designation | POLYBOND-S 112 gsm | ||
Product description | Polymer reinforced web for steam sterilization 112gsm | ||
Recommended sterilization | Steam | ||
STANDARD TEST METHOD | UNIT | TYPICAL VALUE | |
Substance | ISO 536 | g/m² | 112 |
Thickness | ISO 534 | µm | 135 |
Thickness | ISO 534 | mil | 5.3 |
Average tensile strength | ISO 1924-2 | kN/m | 6,85 |
Average wet tensile strength | ISO 3781 | kN/m | 3.15 |
Burst strength | ISO 2758 | kPa | 530 |
Wet burst strength | ISO 3689 | kPa | 260 |
Average initiated tear resistance | ISO 1974 | mN | 1150 |
Cobb test | ISO 535 | g/m² | 18.0 |
Water repellency | EN868-2: app.D | s | 100 |
Pore size | EN868-6 app.E | μm | 15 |
Fluorescence | EN 868-2: 2009 app.B | pts/dm² | 0.0 |
Air permeance | ISO 5636-3 | μm/Pa.s | 5.70 |